TECHNICAL DATA
Calibration
J type TC (Fe/Cu-Ni) - DIN 43710
K type TC (Ni-Cr/Ni-Al) - DIN 43710
Pt100Ω/0°C RTD - DIN 43760 (refer to table 1)
Tolerance:
thermocouple ± 2°C at 200°C
RTD ± 0.55°C at 200°C
Connection:
bi-polar 2 m cable
2 screw terminals (P105 mod.)
TC polarity and colour - DIN 43713, non-polarised RTD (refer to table 2)
Assembly:
TC with grounded junction
RTD with insulated element
Housing material:
stainless steel AISI 304
brass ring (T104-J and T104-K mod.)
Operating temperature:
-100 ... +450°C
-100 ... +600°C (T101-J and T102-K mod.)
Temp.(°C)
TC-J (mV)
TC-K (mV)
Pt100 (. )
-100
-4,63
-3,55
60,25
-50
-2,43
-1,89
80,31
0
0
0
100
50
2,65
2,02
119,4
100
5,37
4,1
138,5
150
8,15
6,14
157,31
200
10,95
8,14
175,84
250
13,75
10,15
194,07
300
16,56
12,21
212,02
350
19,36
14,29
229,67
400
22,16
16,4
247,03
450
25
18,51
264,11
500
27,85
20,64
280,9
Table 1: SENSOR CALIBRATION
Sensor
Positive
Negative
J type TC
red wire
blue wire
K type TC
red wire
green wire
Pt100 RTD
(non-polarised)
Table 2: SENSOR CONNECTION
UNI EN ISO9001